Roche e411

Categories: ,

Description

Feature
cobas e 411 analyzer
System
Fully automated, immunoassay analyzer for random access processing of ECL-based immunoassays (cobas e system format)
Types of modules
  1. cobas e 411 disk analyzer
  2. cobas e 411 rack analyzer
  1. Optional System Table (cabinet); Optional System Table Extension (for printer)
System components
Self contained bench top analyzer comprising an analytical unit and a customized user interface
System interfaces
RS232 serial interface, bidirectional
Throughput
Up to 86 tests/hour
Number of reagent channels
18 channels (reagent slots) for maximum of 18 tests
Programmable test parameters
N/A, programming-by-loading (PBT) concept, application data are transferred without operator intervention from the 2D barcode of the reagent pack (RP) into the instrument database
Sample material
Serum/plasma, urine, others
Sample input/output
  1. Disk Model: 30 positions for samples, calibrators and controls
  2. Rack Model: 15 racks with 5 samples each (= 75 samples in/out)
  3. STAT port: STAT samples are processed with priority
Sample volume
10 – 50 μl
Sample clot detection
Standard (pressure sensor)
Sample barcode types
Code 128; Codabar (NW 7); Interleaved 2 of 5; Code 39
Control unit
Microsoft® Windows® XP-based panel PC
Calibrator/QC input
cobas e system-specific barcoded CalSet vials on disk or racks
Calibration methods
Lot calibration (L-cal); Reagent Pack (RP) calibration (R-Cal)
QC methods
  1. Individual QC and cumulative QC
  2. Up to 100 controls pre-programmable
  1. Preventive QC after calibration of the stand-by cobas e packs
Data storage capacity
  1. The memory contains data files necessary for the analyzer and software to work together:
    • – Reagent Data File: Up to 300 reagent packs
    • – Sample Data File: Up to 2000 test records (for samples and controls)
    • – Calibration Data File: Up to 160 calibrators
    • – QC Data File: Capacity up to 100 controls
    • – Operating Parameter Data File: Up to 305 reagent applications
    • – Up to 20 operator IDs
Electrical requirements
  1. 230/110 Volts AC; 1,000 kVA (disk), 1,250 kVA (rack)
  2. Frequency: 50 Hz or 60 Hz +/- 0.5%
Water/Waste requirements
  1. Water: Bacteria free, de-ionized water supply, resistance of < 10 μS/cm
  2. Liquid waste: Onboard waste container (4 liter), direct drain optional
Operating conditions
  1. Ambient temperature: 18 to 32°C
  2. Ambient humidity: 20 to 80% RH (without condensation)
  3. Heat Output: 2,879 kJ/hr (analyzer unit)
  4. Noise Output: 60 dBA (stand-by), 63 dBA (operation avg.)
Physical dimensions
  1. Width (disk/rack): 120 cm / 170 cm 47.2 in / 67 in
  2. Depth (disk/rack): 73 cm / 95 cm 28.7 in / 37.4 in
  3. Height: 80 cm / 31.4 in (closed top cover) 109 cm / 43 in (opened top cover)
Weight
  1. Disk: 180 kg / 397 lbs
  2. Rack: 250 kg / 551 lbs